Simple measurement of direct current values. Clamp-type logger
[Features]
· Start logging with the touch of a button. Data is logged into the internal memory at predetermined intervals (the default value is 1 second).
· The average, maximum and minimum values from the logged data can be checked on the device.
· High-speed sampling. Minimum sampling time is 500ms.
· Long-lasting battery. Six days of recording is possible on one full charge. *At 1 second intervals (23°C).
· The battery will last approximately 27 days at a recording interval of 1 minute.
· Environment resistant.
· Operating temperature range: -10°C to +55°C.
· Operates even in extreme temperature environments.
Japanese Only
Part Number
Configured Part Number is shown.
ZN-DCC11

ZN-DCC11 appearance
Power supply: secondary lithium battery (USB charging)

ZN-DCC11 dimensional drawing
Unit: mm
| Format | Model ZN-DCC11 |
|---|---|
| Rated Measurement Current | ±200 A DC |
| Maximum Number Of Times Clamped | 10,000 Times |
| Inner Diameter Of Clamp | φ23 |
| Applicable Circuit Voltage | 500 V DC max. |
| Display | 7-segment LCD with 4 digits on 1 row, supplemental information indicators and CHARGE LED |
| Recording Interval | 0.5 s, 1 s, 2 s, 5 s, 10 s, 20 s, 30 s or 1 min (can be set with utility software) |
| Conversion Function *1 | Instantaneous Power |
| Internal Memory Capacity | 105 Files Max. *2 Approx. 1,700,000 data items (1 data item: approx. 4 bytes) *3 |
| Power Supply | Secondary Lithium Battery (charged via USB port on a PC) |
| Current Consumption | 75 mA Max. (maximum current consumption when connected via USB) |
| Battery Life *4 | Approx. 6 days (with recording interval at 1 s and ambient temperature of 23°C) |
| Charging Time | Approx. 3 Hours (estimated time with an ambient temperature of 23°C, starting from a battery level of 0) |
| Operating Temperature Range | Battery powered: -10°C to +55°C (no condensation or freezing) Charging via USB: 0°C to 40°C (no condensation or freezing) |
| Operating Humidity Range | 20 to 80% RH (no condensation or freezing) |
| Storage Temperature And Humidity Range | -10°C to +60°C, 20 to 80% RH (no condensation or freezing) |
| Insulation Resistance | With USB Connector Case: 50 MΩ Min. (500 V DC Mega) |
| Dielectric Withstand Voltage | With USB Connector Case: 1,000 V AC, 50/60 Hz for 1 min |
| Vibration (Resistance) | 10 to 150 Hz, with double amplitude of 0.7 mm, acceleration of 50 m/s2 for 80 min each in X, Y and Z directions |
| Shock (Resistance) | 150 m/s2 3 times each in 6 directions (up/down, left/right and forward/backward) *5 |
| Material | ABS |
| USB Insertion/Removal Durability | 5,000 Times |
| Degree Of Protection | IP30 |
| Weight (Packaged) | Approx. 300 g |
| Accessories | Instruction manual, utility disc (CD-ROM) *6, USB cable (mini-B), and hand strap (comes attached to the logger) |
*1. Instantaneous power is calculated using the measured current and set voltage. Make sure to set the correct number.
*2. A single file is created for each time logging is started/stopped. If the maximum number of files is exceeded, an E12 error will be displayed and logging will be unable to start.
*3. If the number of data items reaches the internal memory capacity, an E11 error will be displayed and logging will stop.
*4. The duration of the battery life depends on the measurement environment, recording interval and operating mode.
*5. Install the product in a location where it will not be subject to impact.
*6. The utility disc contains the utility software, USB driver, PDF manual file, etc.
System requirements for the enclosed software: OS: Windows XP SP3 / Windows 7 (64‑bit supported only for Windows 7)
CPU: Intel-equivalent processor of 1.5 GHz minimum, Memory: 1 GB minimum (2 GB or higher is recommended)
| Format | Model ZN-DCC11 | |
|---|---|---|
| Rated Measurement Current | ±200 A DC | |
| Allowable Measurement Current | 110% of rated measurement current | |
| Current Accuracy* | 0 to ±24.9 A | ±2.0% rdg ±5 digits |
| ±25.0 to ±99.9 A | ±2.5% rdg ±5 digits | |
| ±100 to ±200 A | ±2.0% rdg ±1 digits | |
| Logged Values | Current Values | |
| Applicable Circuit | Direct Current (DC) Circuit | |
*At an ambient temperature of 23±10°C and rated input.
Please refer to the catalog for any product information not listed above.
| Part Number |
|---|
| ZN-DCC11 |
| Part Number | Standard Unit Price | Minimum order quantity | Volume Discount | Days to Ship |
|---|---|---|---|---|
MYR 2,173.30 | 1 Piece(s) | 1 Day(s) |
Loading...
| Type | Clamp power logger | Usage Form | Handy | External storage | Not Applicable |
|---|---|---|---|---|---|
| Battery drive time(h) | 144 | Display | Liquid crystal 7 segment 4 digits 1 levels | Measurement DC Current Range (A) | ±200 |
| Measurement Application Circuit (Power Logger) | DC System Circuit (DC) | Function | Instantaneous Power | Clamp Inner Diameter (mm) | φ23 |
| Material | ABS | Protective Structure | IP30 |

How can we improve?
How can we improve?
While we are not able to respond directly to comments submitted in this form, the information will be reviewed for future improvement.
Customer Privacy Policy
Thank you for your cooperation.
While we are not able to respond directly to comments submitted in this form, the information will be reviewed for future improvement.
Please use the inquiry form.
Customer Privacy Policy